健康人群A1滑车高频超声测值及其影响因素分析

Thickness of A1 Pulley Measured by High-frequency Ultrasound and Its Influence Factors in Healthy Volunteers

  • 摘要:
      目的  探讨健康人群屈指肌腱A1滑车厚度及其影响因素。
      方法  应用高频超声观察90例健康人双侧手指屈指肌腱的10个A1滑车的声像图,并测量其厚度,同时记录每个人的性别、年龄、身高、体质量及体质量指数。分析手指A1滑车厚度的影响因素。
      结果  高频超声能够清晰显示A1滑车结构;双侧手指A1滑车厚度差异无统计学意义(P>0.05)。正常各手指A1滑车厚度差异有统计学意义(P<0.05),进一步比较发现A1滑车厚度可分为两个子集:拇指和小指〔(0.196±0.051) mm〕、食指中指和环指〔(0.230±0.055) mm〕,前者厚度较后者薄;A1滑车厚度与年龄正相关(r=0.468,P<0.001)。拇指与小指在<19岁、20~49岁、≥50岁的正常参考值范围分别为0.09~0.23 mm、0.12~0.30 mm、0.12~0.32 mm;食指、中指及环指在<19岁、20~49岁、≥50岁的正常参考值范围分别为0.11~0.27 mm、0.15~0.35 mm、0.17~0.35 mm。不同性别、不同体质量指数的A1滑车厚度差异无统计学意义(P>0.05)。
      结论  A1滑车厚度的影响因素为部位、年龄。高频超声可以清晰观察及测量A1滑车,成为A1滑车病变定量评价的候选手段。

     

    Abstract:
      Objective  To investigate A1 pulley thickness of flexor tendon in healthy volunteers and to analyze its influence factors.
      Methods  The study included 90 healthy volunteers and the A1 pulley thickness at bilateral fingers was measured using high frequency ultrasound. The following parameters were recorded for each participant: age, gender, weight, height, body mass index (BMI).
      Results  High-frequency ultrasound can clearly show A1 pulley. There was no significant difference in A1 pulley thickness between the bilateral fingers (P>0.05). A1 pulley thickness was significantly different in different fingers (P<0.05). Further comparison showed that A1 pulley thickness could be divided into two subsets: thumb and little finger ((0.196±0.051) mm), index, middle and ring fingers ((0.230±0.055) mm). A1 pulley thickness was positively correlated with age (r=0.468, P<0.001). The normal reference ranges for thumb and little finger were 0.09-0.23 mm, 0.12-0.30 mm and 0.12-0.32 mm, respectively. The normal reference ranges for index, middle and ring fingers were 0.11-0.27 mm, 0.15-0.35 mm and 0.17-0.35 mm in volunteers aged 3-19 yr., 20-49 yr., and ≥50 yr., respectively. Gender and BMI had negligible impact on A1 pulley thickness (P>0.05).
      Conclusion  High-frequency ultrasound can clearly show and measure A1 pulley. Site and age should be taken into account when determining the reference range of normal A1 pulley thickness. High-frequency ultrasound can be a quantitative evaluation method for A1 pulley lesions.

     

/