Objective To measure with standard microbiology methods the sensitivity of 4 commonly used testing methods for Helicobacter pylori (Hp) and to conduct a comparative study of the correlations and differences across the 4 methods.
Methods With the Hp standard strain (SS1) as the reference, colony forming units (CFU) as the units of quantitative analysis for detection performance, and gradient dilution of SS1 suspension as the simulation sample, we measured the sensitivity of 4 Hp testing methods, including bacterial culture, rapid urease test, antigen test, and quantitative fluorescent PCR. CFU values at different concentrations corresponding to the 4 commonly used Hp testing methods were documented and the correlations and differences were analyzed accordingly.
Results The sensitivity of Hp bacterial culture, rapid urease test, antigen test and quantitative fluorescent PCR was 2.0×10 CFU/mL, 2.0×105 CFU/mL, 2.0×105 CFU/mL, and 2.0×102 CFU/mL, respectively.
Conclusion The testing turnover time and sensitivity of different laboratory methods for Hp testing varied significantly. The quantitative fluorescent PCR and bacterial culture both showed relatively high sensitivity, but bacterial culture has complicated operation procedures and is too time-consuming. The rapid urease test and antigen test both were simple and quick to perform, but showed low sensitivity. For clinical and laboratory testing of Hp, appropriate testing method that can identify the corresponding changes of Hp should be selected according to the actual testing purpose.