Welcome to JOURNAL OF SICHUAN UNIVERSITY (MEDICAL SCIENCE EDITION)
XU Xu, PENG Wan-lin, ZHANG Jin-ge, et al. The Application Value of Artificial Intelligence-based Filtering and Interpolated Image Reconstruction Algorithm in Abdominal Magnetic Resonance Image Denoising[J]. Journal of Sichuan University (Medical Sciences), 2021, 52(2): 293-299. DOI: 10.12182/20210360104
Citation: XU Xu, PENG Wan-lin, ZHANG Jin-ge, et al. The Application Value of Artificial Intelligence-based Filtering and Interpolated Image Reconstruction Algorithm in Abdominal Magnetic Resonance Image Denoising[J]. Journal of Sichuan University (Medical Sciences), 2021, 52(2): 293-299. DOI: 10.12182/20210360104

The Application Value of Artificial Intelligence-based Filtering and Interpolated Image Reconstruction Algorithm in Abdominal Magnetic Resonance Image Denoising

  •   Objective  To compare the noise reduction performance of conventional filtering and artificial intelligence-based filtering and interpolation (AIFI) and to explore for optimal parameters of applying AIFI in the noise reduction of abdominal magnetic resonance imaging (MRI).
      Methods  Sixty patients who underwent upper abdominal MRI examination in our hospital were retrospectively included. The raw data of T1-weighted image (T1WI), T2-weighted image (T2WI), and dualecho sequences were reconstructed with two image denoising techniques, conventional filtering and AIFI of different levels of intensity. The difference in objective image quality indicators, peak signal-to-noise ratio (pSNR) and image sharpness, of the different denoising techniques was compared. Two radiologists evaluated the image noise, contrast, sharpness, and overall image quality. Their scores were compared and the interobserver agreement was calculated.
      Results  Compared with the original images, improvement of varying degrees were shown in the pSNR and the sharpness of the images of the three sequences, T1W1, T2W2, and dual echo sequence, after denoising filtering and AIFI were used (all P<0.05). In addition, compared with conventional filtering, the objective quality scores of the reconstructed images were improved when conventional filtering was combined with AIFI reconstruction methods in T1WI sequence, AIFI level≥3 was used in T2WI and echo1 sequence, and AIFI level≥4 was used in echo2 sequence (all P<0.05). The subjective scores given by the two radiologists for the image noise, contrast, sharpness, and overall image quality in each sequence of conventional filtering reconstruction, AIFI reconstruction (except for AIFI level=1), and two-method combination reconstruction were higher than those of the original images (all P<0.05). However, the image contrast scores were reduced for AIFI level=5. There was good interobserver agreement between the two radiologists (all r>0.75, P<0.05). After multidimensional comparison, the optimal parameters of using AIFI technique for noise reduction in abdominal MRI were conventional filtering+AIFI level=3 in the T1WI sequence and AIFI level=4 in the T2WI and dualecho sequences.
      Conclusion  AIFI is superior to filtering in imaging denoising at medium and high levels. It is a promising noise reduction technique. The optimal parameters of using AIFI for abdominal MRI are Filtering+AIFI level=3 in the T1WI sequence and AIFI level=4 in T2WI and dualecho sequences.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return